| USING NARROW-BORE HPLC COLUMNS |
| Number 91061AB |
Solvent consumption is directly related to column volume
Mobile phase solvents add considerable expense to the cost of an HPLC analysis, often more
than the chromatographic column. Not only are the solvents costly to buy, but their
disposal is becoming a considerable expense, as well as an increasingly important
environmental issue.
One way to reduce solvent consumption in HPLC laboratories is to use chromatographic columns with small internal volume. Narrow-bore HPLC columns (2.1 mm ID) have significantly lower column volume than standard HPLC columns and can, therefore, reduce solvent consumption by almost 80%.
Table 1
|
Narrow-bore columns often perform poorly because of excessive
extra column volume.
Reducing solvent consumption in the HPLC laboratory is not just a simple matter of
choosing to use narrow-bore columns. Most HPLC systems in use today are designed for
standard-bore columns and cannot effectively use columns with internal volumes less than
0.5 mL. The reason for this is the extra column volume, the system volume measured from
the injector through the detector (see Figure 1), that reduces the achievable efficiency
of the column. If the calculated value for theoretical plates is less than 90% of the
manufacturer's reported value, there is probably an excessive amount of extra column
volume in the system for the column being used. Also, if early eluting peaks have more
"tailing" than later eluting peaks, there may be a problem with extra column
volume. Early eluting peaks are less diluted and have smaller volumes than late eluting
peaks and will consequently be more susceptible to the detrimental effects of excessive
extra column volume.
FIGURE 1
|
FIGURE 2
|
|||||||||||||||||
Figure 2 provides an example of what can happen when a narrow-bore column is used with a typical HPLC system. The narrow-bore column is unable to achieve the baseline separation provided by the standard bore column. However, when the narrow-bore column is used in an "optimized" HPLC system that has less than 10 microliters of extra column volume, excellent resolution and peak shape is obtained.
To obtain the best performance from narrow-bore columns, extra column volume
must be minimized.
If excessive extra column volume is a problem, several things can be done to
improve the performance of the system.
Table 2
|
Narrow-bore columns normally require detector flow cells with less than 2 microliter volumes. The design of the flow cell, however, can be just as important as the cell volume. Some low volume flow cells perform worse than larger volume flow cells because of their inadequate design.
![]()
MAC-MOD Analytical, Inc. -- info@mac-mod.com -- 1-800-441-7508
![]()