| Section
3 What is the Best Column |
|
![]()
Will a Shorter Column Work for Your Application?
It is not necessary to actually install, equilibrate and run a sample on a shorter column to determine if it will provide sufficient resolution for your sample mix. if you already have data from a longer column, a few quick calculations will tell you if a shorter column having the same particle size will work.
| Estimating
Change in Rs Resolution, Rs, Rs2 = Rs1 x (L2/L1)0.5 Where: |
Estimating
Changes in Analysis Time and Back Pressure In similar fashion, T2
= T1 x L2/L1 Where: |
An Example
Figure 3-1 Column: 4.6 x 150 mm; Pressure: 1050 psi
|
Suppose that you are using a 25 cm column. The resolution between the least resolved peak pair is 3.7, the total analysis time is 18 minutes and the back pressure is 1050 psi. If resolution greater than 2.5 is required for this analysis, could a 15 cm column be used in place of the 25 cm column? How would the analysis time and pressure change?
Rs2 = 3.7 x (15/25)0.5 = 2.9
T2 =18 min x 15/25 = 10.8 min
P2 =1050 psi x 15/25 = 630 psi
Where:
Rs2 = Resolution with 15 cm column
T2 = Analysis time with 15 cm column
P2 = Pressure with 15 cm column
It is clear from the analysis that a 15 cm column could be substituted and perform satisfactorily. This change would result in using 40% less solvent and reducing analysis time by 7 minutes.
| The example above shows that you can quickly determine if a shorter column will work for a particular analysis without having to actually install and run the column. This can save you many hours of trial-and-error work in the laboratory. |
| For Technical Assistance Call |
|
| 1-800-441-7508 |
Section 3 | Section 3: Column Diameter | Table of Contents | Next Section
![]()
MAC-MOD Analytical, Inc. -- info@mac-mod.com -- 1-800-441-7508
![]()