| Section
6 Injection-to-Injection Reproducibility |
|
Retention and Resolution Variability
If peak retention
(tr) and
resolution (Rs) are not reproducible from injection-to-injection, or even
day-to-day, the accuracy and precision of the HPLC method may not be acceptable. Poor
reproducibility is usually caused by either changes related to chromatographic conditions
or variability in column performance (see Table 6-1).
Table 6-1
|
|||||||||||||||||||||||
Identifying and Correcting Variability in Column
Performance
Reproducibility problems
caused by
variability in column performance usually show up as continuous changes in retention time
and/or loss in peak symmetry. Identifying and correcting the cause of column variability
can be difficult, however, a systematic approach to taking corrective action will usually
be successful. Table 6-2 lists the causes of poor injection-to-injection reproducibility
that are due to variability in column performance along with identifying symptoms and
recommended corrective actions.
|
||||||||||||||||||||||||
This general troubleshooting information for injection-to-injection reproducibility problems applies to isocratic and gradient HPLC methods; but gradient methods require some additional troubleshooting hints.
Retention Time Variability in Gradient Separations
Gradient separations
have many advantages
for the analysis of complex samples with components that vary widely in polarity. During a
gradient separation, the column and analytes are subjected to a constant change in mobile
phase conditions. This makes troubleshooting column or system problems more difficult. For
example, the problem of gradient elution retention time irreproducibility, varying
retention times from injection-to-injection, has several unique causes.
| CAUSE: Incomplete column equilibration | |||
| SOLUTION: Make sure the column is equilibrated | |||
| A standard 4.6 x 250 mm column has an internal column volume of
approximately 2.5 mL and needs at least 10 to 15 column volumes to completely
equilibrate.Your gradient equilibration time may need to be adjusted to match this volume
according to your flow rate. Duplicate injections after equilibration will verify adequate
equilibration. Column equilibration is easier when both the organic and aqueous/buffer components of a mobile phase contain the same concentration of mobile phase additives. For instance, TFA is often added to a mobile phase to improve peak shape, especially for peptide separations, and should be added to both the organic and aqueous mobile phase components. |
|||
| CAUSE: Highly retained material "fouls" the column | |||
Not all sample material may be completely eluted from a column if the
final gradient composition is not strong enough. For example, the analysis of a sample in
a complex matrix may require a gradient that starts at 10% methanol and ends at 60%, but
it may require 75%-100% organic to elute all matrix components. Figure 6-1 shows peaks
eluting late in a blank gradient run of 0-1000% methanol. This example demonstrates that
the gradient conditions of 0-60% methanol are not sufficient to remove all matrix
components. If the organic composition is not increased beyond 75%, matrix components will
build up on this column and the result may be changes in retention time of eluting sample
peaks. Noneluted material can be removed by extending both the gradient % organic and
gradient time proportionately or washing the column with 100% methanol after a
to-be-determined number of sample injections.
|
|||
| SOLUTION: Back-flush the column with a strong solvent |
|||
| If the column has become heavily contaminated, first back flush the column. Disconnect the column from the detector, turn it in the opposite flow direction and Bush out all buffers with an aqueous/ organic mobile phase. Then, switch to 100% organic. Acetonitrile is a good organic solvent to start with, since it is stronger than methanol in a reversed-phase system and it may elute additional strongly retained components. If this flush is not sufficient, isopropanol is another good solvent choice. Isopropanol is more effective at solubilizing many proteins, peptides, and fats and is a stronger solvent than acetonitrile or THF. Unlike THF, isopropanol is compatible with PEEK tubing. |
| For Technical Assistance Call |
|
| 1-800-441-7508 |
Previous Section | Table of Contents | Next Section
![]()
MAC-MOD Analytical, Inc. -- info@mac-mod.com -- 1-800-441-7508
![]()