How to Regenerate Heavily Fouled Columns
| Symptoms of Column Fouling Since many HPLC Some of the symptoms of column fouling are |
| |
increases in back pressure | |
| |
changes in retention times | |
| |
peak broadening or tailing and | |
| |
loss of resolution |
Column Regeneration
The performance
of a fouled, reversed-phase
column can often be restored by purging the column with a strong eluting solvent such as
100% acetonitrile. If this does not restore column performance follow the column
regeneration procedure described on the next page before opening or discarding the column.
A procedure to regenerate normal phase columns is also included.
| Regeneration Procedure for Reversed-Phase Columns
Note: |
Regeneration Procedure for Normal-Phase Columns
Caution |