Number 00511TR

Maximizing HPLC Reproducibility in Highly Aqueous Mobile Phases

Poor Retention Time Reproducibility is a Common Problem When Operating With Highly Aqueous Mobile Phases
When separating very polar, water-soluble compounds, it is not unusual to use mobile phases that contain less than 10% organic modifier (CH3OH, CH3CN, etc.) in order to achieve sufficient retention. However, operating under such highly aqueous mobile phase conditions can lead to poor chromato-graphic reproducibility. Over time peaks will elute with shorter and shorter retention times and resolution between peaks will deteriorate (Figure 1). If the column is allowed to stand (no flow) in a highly aqueous mobile phase, a dramatic change in peak retention time can occur making the column unusable (Figure 2).

FIGURE 1
Poor Reproducibility Due to Phase Collapse

  Column: YMC ODS
Mobile Phase:
5% CH
3OH
95% 0.1 M KH
2PO4           
Flow Rate: 1 mL/min
Sample:
1. Vitamin C
2. Vitamin B1
3. Vitamin B6 
4. Nicotinamide

 
 After only 72 hours of operation, the retention time for
nicotinamide decreased by over 8% due to phase collapse.

 

FIGURE 2
Phase Collapse Reduces Retention and Degrades Resolution

Column: 
Mobile Phase:
Zorbax StableBond SB-C18, 4.6 x 150 mm
10% Acetonitrile
90% 0.050 M Phosphate buffer, pH 2.6
 
  Sample:   1. Uracil
2. Nitroethane
3. Phthalic acid
4. 4-chloroaniline
5. 3-cyanobenzoic acid
6. 3,5-dimethylaniline
7. 1-nitrobutane
 
Letting a C18 or C8 column stand in a highly aqueous mobile phase will accelerate phase collapse. In this case, a C18 column was stored overnight in 100% water. The next day chromatogram B was generated under the identical conditions as chromatogram A. Chromatogram B shows the dramatic reduction in retention times and resolution that can occur because of phase collapse.

Phase Collapse Causes Peaks to Elute With Less Retention
This change in retention time is caused by what many researchers call “phase collapse,” or “matting,” of the hydrophobic C18 or C8 alkyl phases under highly aqueous mobile phase conditions (Figure 3). As phase collapse progresses, the availability of the alkyl phase to interact with solutes decreases and retention time decreases. There is also evidence that this phase collapse can even displace aqueous mobile phase from the pores of the stationary phase, reducing the surface area accessible to solutes and further reducing retention times.  

FIGURE 3
Phase Collapse

Under normal conditions the alkyl phase is extended into the mobile phase and solvent and sample molecules have full access to the stationary phase. When using highly aqueous mobile phases, the alkyl phase has a tendency to collapse on itself. When this happens, retention times will be affected.

Figure 4 provides an example of this effect. After letting this base deactivated C8 column stand for only 10 minutes in a 100% aqueous mobile phase, retention time decreases by over five minutes. Figure 4 also shows that purging with organic solvent can restore retention on the column. Organic solvent “solvates” the hydrophobic bonded phase, causes the alkyl phase to once again extend into the mobile phase, and allows solutes access to the surface area inside the pores. Unfortunately, if a highly aqueous mobile phase continues to be used, the phase may begin to collapse again and retention times will change.

FIGURE 4
The Effect of Phase Collapse on Retention

Initial Injection:

 

After standing for 10 minutes in mobile phase 8.58 min.:

 

After purging with organic solvent:

 

Mobile Phase:    
Columns: 
Sample:
0.1% Acetic acid in water
Base deactivated
Amoxicillin

After letting this C8 column stand for only 10 minutes in 100% aqueous mobile phase, the retention time for amoxicillin drops by over five minutes, indicating phase collapse. By purging the column with organic solvent, the alkyl phase is again extended into the mobile phase and retention time is restored.

Maximizing HPLC Reproducibility When Using Highly Aqueous Mobile Phases
If you are experiencing a problem with retention time reproducibility while using mobile phases that contain less than 10% organic modifiers, consider one of the following corrective actions:

  1. Purge the column periodically with a mobile phase containing more than 50% organic modifier. Each situation is different, but if retention times drop by more than 5%, it is probably time to purge the column.
  2. Don't let a highly aqueous mobile phase stand in your column. This will avoid promoting phase collapse and the associated displacement of aqueous mobile phase from the stationary phase pores.
  3. If the column shows poor retention as a consequence of having been left standing in a highly aqueous mobile phase, condition the column by purging with a mobile phase containing at least 50% organic modifier. In some cases, you may have to purge with a mobile phase containing more than 75% organic modifier. It also helps to purge at higher pressure to force mobile phase into the pores.
  4. Consider using a column that does not exhibit problems with phase collapse.

Some HPLC Columns Do Not Exhibit Problems With Phase Collapse
Column manufacturers have addressed the problem of phase collapse by using polar silanes as the bonded phase, or by using hydrophilic end-capping (Figure 5). Both approaches have the same desired effect of keeping the alkyl phase extended in the mobile phase, even when using 100% aqueous mobile phase. By keeping the alkyl phase extended, chromatographic reproducibility is greatly enhanced.

FIGURE 5
Polar Embedded Phase and Hydrophilic
End-Capped Stationary Phases

Phase collapse can be prevented by embedding polar groups into
the alkyl phase or by using hydrophilic end-capping.

 

Table 1
HPLC Columns Designed Specifically
for Highly Aqueous Mobile Phase

AquaSep
HydroBond AQ
HydroBond PS
ProntoSIL AQ
YMC ODS-AQ
Zorbax SB-Aq

 AQ Type Phases
Some commercial columns are marketed specifically for highly aqueous mobile phase conditions (Table 1 above). These columns offer the advantage of improved retention for very water-soluble com- pounds (Figure 6) as well as improved reproducibility. Retention is usually sufficient on these phases to avoid having to use ion pair reagents, therefore, chromatographic conditions are simpler and methods are more rugged.

FIGURE 6
Improved Retention for Highly Water-Soluble Compounds


 

Column:
HydroBond AQ C8 4.6 x 250 mm, 5 µm      
Mobile Phase:
0.05M KH
2PO4, pH 2.4
Flow Rate:
1.0 mL/min
Sample:
1. Oxalic acid
2. Tartaric acid
3. Formic acid
4. Malic acid
5. Malonic acid
6. Lactic acid
7. Maleic acid
8. Fumaric acid

Polar embedded phases or stationary phases with hydrophilic end-capping usually provide improved retention for highly water soluble compounds.

Polar Embedded Phases
Most polar embedded phases are marketed as specialty columns and positioned by manufacturers as offering enhanced polar selectivity (Figure 7) and improved peak shape for basic compounds (Figure 8). However, these polar embedded phases also have the advantage of operating with highly aqueous mobile phases without experiencing problems with phase collapse. Table 2 provides a list of polar embedded phase columns that we have found to work well under high aqueous conditions. 

Figure 7
Some Polar Embedded Phases Offer Enhanced Polar Selectivity

Mobile Phase:
65% Acetonitrile
35% 0.050 M Phosphate buffer,       
pH 3.2
Sample:
1. Uracil
2. Pyridine
3. Phenol
4. N,N-Dimethylaniline
5. p-Butylbenzoic acid
6. Toluene

The ProntoSIL C18-EPS is a polar embedded phase that exhibits greater selectivity for this mixture of polar compounds than a typical base deactivated phase.

 

Table 2
Polar Embedded Phases That Do Not
Exhibit Problems With Phase Collapse

Discovery RP-Amide C16
Hypersil HyPURITY Advance
Keystone Prism
ProntoSIL C18-EPS
Symmetry Shield
Zorbax Bonus-RP
       

Figure 8
Polar Embedded Phases Provide
Improved Peak Shape

Mobile phase: 80% Methanol,
20% 0.025 M Phosphate buffer,
pH 6.0
Analyte: Amitriptyline

This polar embedded phase (ProntoSIL C18-EPS) shows improved peak shape for basic compounds. Its embedded amide group shields solutes from interacting with silanols on the silica stationary phase support and, thereby, minimizes peak tailing.

Conclusions
Typical C18 and C8 columns often suffer from poor reproducibility when used with highly aqueous mobile phases. Avoid mobile phases that contain less than 10% organic modifier when using these columns. If you are not able to achieve adequate retention for your sample without using a highly aqueous mobile phase, we recommend using a column that is immune to phase collapse. Tables 1 and 2 above provide a list of these types of columns. If your separation suffers from phase collapse and you are not able to substitute one of the columns in Figures 1 or 2, be sure to purge your column frequently with a mobile phase containing at least 50% organic modifier to reduce the effect of phase collapse. 

Suggested HPLC Columns for
Highly Aqueous Mobile Phase Conditions

 


MAC-MOD Analytical, Inc. -- info@mac-mod.com -- 1-800-441-7508